Detail LUTRON EMF-819
3 Axis RF Electromagnetic Field Meter
- 3 axis probe
- Radio frequency ellectromagnetic field tester
- Wide measuring frequency ranges, 50 MHz to 3 GHz
- EMF-189 is used for broadband devices of monitoring the wide range radio frequency electromagnetic field value
- For precision measurement consideration, the meter is included one probe : EP-05H ( Frequency range : 50 MHz to 3 GHz )
- Unit : V/m, W/m W/m^2, mW/cm^2
- Frequency selection, two points : Normal, 2.45 GHz
- Alarm setting function can warn the user if the measuring antenna is to near the strong radiation sources, the buzzer will sound to remind the user
- Peak hold function to latch peak value
- Data hold function to lock the current reading
- RS232 computer interface
- Hard carrying case is included
- Large size LCD with contrast adjustment, which can fit best viewing angle
- Microcomputer circuit provides special function and offers high accuracy
- Powered by 006P DC 9V battery or DC 9V adapter
APPLICATIONS :
This meter is specially developped for measuring or monitoring electromagnetic field, for example : Cell-phone station, hospital equipment, radar, micro-wave oven, radiation work, TV antenna, Radio station, welding equipment, baking-equipment, television, computer, factory, laboratory, and other environment… etc
SAFETY INSTRUCTIONS
DANGER
- For worker’s safety, be aware that persons with electromagnetic implant ( e.g cardiac-pacemarker ) are subject to especial danger in some case
- Particular to onserve the local safety regulations of the operator of the equipment
- Before using the device, it need to know that how to setting ” alarm-limit” value
ATTENTIONS :
- Claims by some scientist that long term exposure to electromagnetic field may be the cause of childhood leukimia and other forms of cancer
- Complete answer to any of these and related questions are not currently available. At the present time the most common practice is to avoid excess exposure over long priod of time
- Complete answer to any of these and related ” prudent avaidance ” as stated by the environmental protection Agency ( EPA ) USA is recomended
- According to ICNIRP of reference levels to time-varying electromagnetic files. The E-field strength levels are :
GENERAL PUBLIC :
Frequency range | e-field strength (V/m) |
10 to 400 MHz | 28 |
400 to 2000 MHz | 1.375 x f^1/2 |
2 to 300 GHz | 61 |
Occuptional :
Frequency range | e-field strength (V/m) |
10 to 400 MHz | 61 |
400 to 2000 MHz | 2 x f^1/2 |
2 to 300 GHz | 137 |
SPECIFICATIONS :
- Circuit : Custom one-chip microprocessor LSI circuit
- Display : LCD size 58×34 mm
- Measurement unit : V/m, W/m^2, mW/cm^2
- Accuracy : < 2 dB
- Probe structure : 3 axis
- Probe input mpedance : 50 OHM
- Frequency selection : Two points selection : Normal, 2.45 GHz
- Sensor structure : Semiconductor
- Data hold : Freeze the display reading
- REC function : Record maximum and minimum value
- Power off : Auto shut off saves battery life or manual off by push button
* Can default auto power off or manual power off
* When default auto power off, power will off automatically after 10 min, if no button be pressed - Peak hold : To latch the peak measurement value
- Alarm setting : Buzzer will sound when display over the setting value
- Sampling time : Approx. 1 second
- Low battery indicator : When display shoe low battery indicator, it should change the batteries.
- Data output : RS 232 PC serial interface
- Operating temp : 0 to 50 ◦C
- Operating humidity : Less than 80% R.H
- Power supply : 006P DC 9V battery ( heavy duty or alkaline type ) OR dc 9v adapter
- Power current : Approx DC 5.59 mA
- Weight : 425g/0.94 LB
- Dimension :
Main instrument : 200.0 x 76.2 x 36.8 mm
Probe : 70 mm ( diameter ) x 240 mm ( length )