Detail LUTRON dt-2269 Digital Stroboscope :
- DT-2269 is a microprocessor circuit design, high accuracy, digital readout STROBOSCOPE, Adjusting the Flash Rate by push button keyboard, unique design in the world, easy operating and with intelligent function. That is ideal for inspecting and measuring the speed of moving gears, fans, centrifuges, pumps, otors, and other equipment used in general industrial maintenance, production, quality control, laboratories and as well as for schools and colleges for demonstrating strobe action.
SPECIFICATIONS :
- Display : 14 mm ( 0.56 inch ) LED, 6 digits
- Set up unit : Flash rate-RPM /FPM, Hz
- Flash rate set up range :
RPM : 5 to 12,500 RPM
Hz : 0.083 to 175 Hz - Flash rate set up range :
RPM : 5 to 12,500 RPM
Hz : 0.083 to 175 Hz - Resolution ( Internal trigger ) :
– RPM/FPM : 0.1 RPM ( 5 to 999.9 RPM ) ; 1 RPM ( 1.000 TO 9.999 RPM ); 10 RPM ( 10.000 to 12.500 RPM )
– Hz : 0.001 Hz ( <10 Hz ); 0.01 Hz ( 10 Hz – 99.99 Hz ); 0.1 Hz ( 100-208 Hz ) - Accuracy :
± (0.15% + 0.2 RPM) rdg : 1000 RPM
± (0.5% + 0.1 RPM) rdg : 1000 to 3.300 RPM
± 1% FS : 3.301 to 12.500 RPM - Switch select function : RPM, Hz, Fine adjust, Coarse adjust, Multiply by 2, Divide by 2, Fast Finder, Memory recall.
- Memory : Can memorize 10 sets of measuring data
- Set p stability : 1 digit within 10 minute
- Data output : RS 232 computer interface
- Operating temp : 0 to 50 ◦C (32 to 122 ◦F )
- Operating humidity : Less than 80% R.H
- External trigger :
Input signal : 5 V to 30 V rms
5 to 10,500 RPM/ 0.083 to 175 Hz - Power supply :
AC110V 10%, 50/60 Hz
AC220V 10%, 50/60 Hz
AC230V 10%, 50/60 Hz - Circuit : This stroboscope/tachometer employs an custom one-chip of microcomputer LSI circuit and crystal control time base which results in extraordinary accuracy and high set up stability over a wide, dynamic range.
- Power consumption : Less than 30 watt
- Dimension : 21x12x112 cm
- Weight : 1 kg/ 2.2 LB
- Housing case : compact and impact plastic injection case with plastic mirror type reflector
- Calibration : Crystal time base and microprocessor circuit, not necessary tak nay calibration procedure if the stroboscope working properly.