Pressure Gauge Model PG43SA-S, NS 100


DESCRIPTION

The model PG43SA diaphragm pressure gauge has been specifically designed for the requirements of sanitary applications.

The purely mechanical pressure transmission, using the diaphragm-element measuring principle, and the high overpressure safety ensure a safe pressure measurement. As a result of the dry measuring cell, the risk of contamination of the product by transmission fluid is eliminated. The flush welded diaphragm element in conjuction with the aseptic process connections enables a dead-space free connections to the process.

The instrument finds applications, for example, in system for ultrapure steam generation or as an independent pressure display, without the need for an external power supply, on portable tanks.

The PG43SA-S in hygienic design can be used for CIP (cleaning in place) and SIP (sterilisation in place) and in wash-down areas, and optionally it canbe copletely sterilised in an autoclave. The measuring instrument can thus be cleaned reliably and time-efficiently. At an easily accessible point on the top of the case, the zero point can be easily corrected.

Based on a third party verification, the model PG43SA-S corresponds to the 3-A Sanitary Standard.

A variety of 3.1 and 2.2 certificates available for GMP-compliant documentation, such as material certificate or the list of single measured values.

APPLICATIONS

  • Hygienic pressure measurement is sanitary applications for the pharmaceutical, biotechnology, food and beverage industries.
  • Mechanical pressure display on pipilines, fermenters, bioreactors and vessels.
  • Pressure/vacuum monitoring during cleaning, sterilisation, pressure testing
  • For gases, compressed air, vapour; liquid, pasty,powdery and crystallising media.

SPECIAL FEATURES

  • Safety via mechanical pressure transmission
  • Case and wetted parts in hygienic design
  • Suitable for SIP and CIP, optionally completely autoclave
  • Simple zero point setting
  • High overpressure safety

SPECIFICATION PRODUCT

Version

EN 837-3

Nominal size in mm

100

Accuracy class

1.6

Pressure limitation

Steady: Full scale value

Fluctuating:

Scale range <4 bar: Full scale value

Scale range ≥4 bar: 2/3 x full scale value

Option: Vacuum safety

Overpressure safety

2x full scale value, max. 40 bar or max. pressure rating (PN) of the process connection

Option: 5x full scale value, max. 40 bar or max. pressure rating (PN) of the process

connection

Permissible temperature

Ambient: -20 … +60 °C

Medium: -20 … +150 °C

CIP and SIP: 150 °C continuously for wetted parts

Option: Autoclavable, max. 134 °C, ≤ 20 minutes

Temperature effect

When the temperature of the measuring system deviates from the reference

temperature (+20 °C): max. ±0.5 % / 10 K of full scale value

Have any Question or Comment?

Leave a Reply

Your email address will not be published. Required fields are marked *

Categories

LOKASI

Copy Protected by Chetan's WP-Copyprotect.